BET surface area is the internal surface of activated carbon in square metres per gram, calculated from how much nitrogen the carbon adsorbs at −196 °C. Coconut shell activated carbon typically falls around 900–1,300 m²/g. Because the test needs specialised equipment, most trade specifications use iodine number as a practical indicator instead.

How the test works

A small sample is degassed under vacuum and heat to remove adsorbed water and gases, then cooled in liquid nitrogen (−196 °C, 77 K). Nitrogen is admitted in steps and the amount adsorbed at each pressure is recorded. The Brunauer–Emmett–Teller (BET) equation is applied to part of this isotherm to estimate how much nitrogen forms a single layer on the surface, and from that the surface area. The procedure is standardised in ISO 9277.

For microporous materials such as activated carbon, the BET result depends on the pressure range chosen for the calculation, so results from different laboratories are only comparable if they use the same method and range. Ask the laboratory to report the range used.

What else the nitrogen isotherm tells you

The same measurement can give total pore volume and a pore size distribution. This is often more useful than the single surface area figure: two carbons with the same BET area can perform differently if one has more pores in the size range of the molecule you need to remove.

BET and iodine number

Iodine number (ASTM D4607) is measured with simple wet chemistry and correlates with micropore volume. As a rough rule of thumb, the iodine number of a carbon in mg/g is of the same order as its BET surface area in m²/g, but this is not a conversion and should not be used to state a BET value that was never measured. See iodine number and hardness explained.

When to ask for BET

For routine purchasing of coconut shell GAC, iodine number, hardness, ash, moisture and particle size are the parameters normally written into contracts. We have not yet published a BET result for our product; it can be measured by an outside laboratory on request.

Test it yourself

For the methods behind every parameter and our independent results, see Quality Control & Laboratory Testing and the full reports on our certification page. To check our carbon in your own laboratory, request a sample with its certificate of analysis.